The 5700A and 5720A are five-function calibrators designed to address the most demanding electrical calibration workload. They source direct and alternating voltage and current and resistance. A wideband voltage option extends overall ac bandwidth to 10 Hz to 30 MHz to cover to RF voltmeters. Both are compatible with the 5725A, and 5205A/5215A amplifiers.

 

SKU: FLU5700A Category: Tag:
Brand:

Description

Voltage DC
Ranges: 0 to ±1100V
5700A Best 1 Year Specification: ±7.0 ppm of setting
5720A Best 1 Year Specification: ±3.5 ppm of setting
Voltage AC
Ranges: 220 mV to 1100V
10 Hz to 1 MHz
5700A Best 1 Year Specification: ±75.0 ppm of setting
5720A Best 1 Year Specification: ±45.0 ppm of setting
Resistance
Ranges: 0 to 100 MO, 18 values in x1 and x1.9
5700A Best 1 Year Specification: ±12.0 ppm
5720A Best 1 Year Specification: ±8.5 ppm
Current DC
Ranges: 0 to ±2.2A (11A with 5725A)
5700A Best 1 Year Specification: ±50 ppm of setting
5720A Best 1 Year Specification: ±35 ppm of setting
Current AC
Ranges: 9 m A to 2.2A (11A with 5725A) 10 Hz to 10 kHz
5700A Best 1 Year Specification: ±140 ppm of setting
5720A Best 1 Year Specification: ±120 ppm of setting
Wideband AC Voltage Option
Ranges: 300 m V to 3.5V
10 Hz to 30 MHz
5700A Best 1 Year Specification: ±0.4% of setting
5720A Best 1 Year Specification: ±0.4% of setting

 

General Specifications
Warm up Time 2x the time since last warmed up, to a maximum of 30 minutes
System Installation Rear output configuration and rack-mount kit available
Interfaces IEEE-488, RS-232, 5725A, 5205A or 5215A, 5220A, phase lock in (BNC), phase reference out (BNC)
Temperature Performance Operating: 0°C to 50°C
Calibration: 15°C to 35°C
Storage: -40°C to 75°C
Relative Humidity Operating: < 80% to 30°C, < 70% to 40°C, < 40% to 50°C
Storage: < 95%, non-condensing. A power-on stabilization period of four days may be required after extended storage at high temperature and humidity.
Safety Designed to comply with UL311; IEC 348-1978; IEC 66E (CO)4; CSA 556B
Guard isolation 20 volts
EMI/RFI Designed to comply with FCC Rules Part 15, Subpart J, Class B; VDE 0871, Class B
Line Power 47 to 63 Hz; + 10% allowed about selectable nominal line voltage: 100V, 110V, 115V, 120V, 200V, 220V, 230V, 240V
————–
Maximum power: 5700A/5720A: 300 VA;5725A, 750 VA
Size 5700A/5720A: Height 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet; width 43.2 cm (17 in), standard rack width; depth 63.0 cm (24.8 in) overall; 57.8 cm (22.7 in), rack depth

5725A:Height 13.3 cm (5.25 in); width and depth same as 5700A/5720A
Both units project 5.2 cm, (2 in) from rack front

Weight 5700A/5720A:27 kg (62 lbs.)
5725A: 32 kg (70 lbs.)

 

Accessories

Part Number Description
Qty
601622 Operator’s Manual
1
1668111 Getting Started Guide
1
601655 Remote Programming Reference Guide
1
601630 Service Guide
1
Service Manual Supplement 8/03
1
601648 Operator’s Guide
1
Power Cord
1

Options

Option Description
03 Wideband AC Voltage Module

Additional information

Weight 62 lbs
Dimensions 7 × 18 × 26 in
Condition

Model Number

5700A Option 006 – Fluke Calibration

Manual

Fluke 5700A Operators Manual https://4gte.com/index.php/filemanager/datasheet/download/id/469/

Manufacturer

Fluke

Type

Content missing

You may also like…