Rohde & Schwarz ZV-Z235 Calibration Kit, 0 Hz – 26.5 GHz, 3.5 mm


General Inquiries/Quote Request
Rental/Lease Information Request

The specified effective system data are established after performing a suitable system error calibration, e.g. TOSM, at a R&S ZVA,
R&S ZVB, or R&S ZVT vector network analyzer, using the characteristic data of the calibration kit, which are stored on a provided
USB stick. This data is valid between +18 ยฐC and +28 ยฐC, at a measurement bandwidth of 10 Hz, and a nominal power of โˆ’10 dBm at
the calibration ports. The calibration kit is fully functional down to 0 Hz, with effective system data as specified below, although the
data is only verified for frequencies as stated.

Directivity

10 MHz to 700 MHz > 36 dB, typ. 46 dB
700 MHz to 24 GHz > 40 dB, typ. 46 dB

Source match

10 MHz to 700 MHz > 30 dB, typ. 43 dB
700 MHz to 24 GHz > 36 dB, typ. 43 dB

Reflection tracking

10 MHz to 700 MHz < 0.2 dB, typ. 0.04 dB 700 MHz to 24 GHz < 0.1 dB, typ. 0.02 dB Load match 10 MHz to 700 MHz > 36 dB, typ. 46 dB
700 MHz to 24 GHz > 40 dB, typ. 46 dB

Transmission tracking

10 MHz to 700 MHz < 0.2 dB, typ. 0.04 dB 700 MHz to 24 GHz < 0.1 dB, typ. 0.02 dB

SKU: R&SZV-Z235 Categories: , Tag:
Brand:

Additional information

Weight 24 lbs

You may also like…